The repair service ensures that the functionality of the GM901 process gas analyzer is restored quickly. The device is also given a general overhaul in order to prevent future failures.

At a glance

About service

The repair service ensures that the functionality of the GM901 process gas analyzer is restored quickly. The device is also given a general overhaul in order to prevent future failures as far as possible. The lump sum for repairs includes preventative maintenance and the replacement of function-related assemblies, consumables, and wear parts. For this reason, the warranty which promises faultless operation is extended to the complete repaired device. The lump sum for repairs speeds up processing and makes it possible to budget for repairs.

Description
  • Reason for failure is identified and repairs are carried out at a fixed price
  • Functionality of the device is restored
  • Measurement certainty and calibration checked under laboratory conditions
  • Firmware update in accordance with customer requirements
  • Preventative replacement of assemblies, wear parts, and consumables
  • Brief repair report
Benefits
  • Fast and reliable repairs and testing performed by us
  • More cost-effective to repair existing components than to install new or exchange units
  • Extends the product service life
  • Avoid systematic faults with advice about preventative measures (optional)
  • No stock changes: original equipment with the same serial number and specific designations after the repair has been completed
  • Definition of standard repair times ensures the measuring device is available again quickly
  • Warranty extended for the complete repaired device

Service Features

Product Family Group

Repair Services

Short description

The GM901 repair service ensures that the functionality of the measuring device is restored quickly.

Application

GM901 process gas analyzer

Prerequisites
  • Device already cleaned
  • Return delivery form completed
  • Non risk decleration form completed
  • Additional information about the application
  • Repair order approved
Scope of Service
  • Function check
  • Fault diagnosis
  • Checking firmware is up to date, updating if necessary
  • Checking the electrical parts
  • Replacement of defective mechanical housing parts
  • Endurance test in climatic cabinet over at least one climate cycle, with data evaluation
  • Operational check by means of system test
  • Final adjustment on customer's measuring path with linearity test and determination of coefficients
Documentation

Brief repair report